Reset Search
 

 

Article

Operating Pressures for CYCLE-TAINER Solvent Delivery Systems

« Go Back

Information

 
Question
What are the recommended operating pressures for the various sizes of CYCLE-TAINER solvent delivery systems?
Answer

Container operating pressures are based on your application and the configuration of your solvent dispensing area.  Below is a summary of the recommended operating pressure ranges for each container type.

 

4 L Model S

18 L Model S

52 L ASME Model S

52 L Model C

52 L non-ASME Model S

215 L ASME Model S

215 L ASME Model C

215 L non-ASME Model S

225 L Model C

1250 L Model S(1)

Operating Pressure

0-15 psi

0-15 psi

0-40 psi

0-15 psi

0-15 psi

0-40 psi

0-40 psi

0-15 psi

0-15 psi

0-19 psi

Pressure Relief Valve venting pressure

50 psi

50 psi

50 psi

50 psi

50 psi

50 psi

50 psi

50 psi

50 psi

25 psi